Devices and pressure sources for administration of nasal continuous positive airway pressure (NCPAP) in preterm neonates

After weaning preterm babies from a ventilator, short binasal prong devices for NCPAP (nasal continuous positive airways pressure) are more effective than single prong devices. Nasal continuous positive airway pressure (NCPAP) is a form of breathing support that is less invasive than mechanical ventilation (where a tube goes down into a baby's lungs). NCPAP usually delivers oxygen to a baby through tubes into the nose, or less commonly, through face masks. It can be used after weaning a baby from ventilation (extubation), or to help babies who need help for lung problems, but do not need ventilation. The review of trials found that short binasal prongs (entering both nostrils) are better than single prong NCPAP for preterm babies. More research is needed on the best pressure delivery system and the best pressure levels to use.

Wprowadzenie

Nasal continuous positive airway pressure (NCPAP) is used to support preterm infants recently extubated, those experiencing significant apnoea of prematurity and those with respiratory distress soon after birth as an alternative to intubation and ventilation. This review focuses exclusively on identifying the most effective pressure source and interface for NCPAP delivery in preterm infants.

Cele

To determine which technique of pressure generation and which type of nasal interface for NCPAP delivery most effectively reduces the need for additional respiratory support in preterm infants extubated to NCPAP following intermittent positive pressure ventilation (IPPV) for respiratory distress syndrome (RDS) or in those treated with NCPAP soon after birth.

Metody wyszukiwania badań

The strategy included searches of MEDLINE (1966 - 2006), the Cochrane Central Register of Controlled Trials (CENTRAL, The Cochrane Library, Issue 4, 2006) CINAHL, abstracts from conference proceedings, cross-referencing of previous reviews and the use of expert informants.

Kryteria doboru badań

Randomised or quasi-randomised trials comparing different techniques of NCPAP pressure generation and/or nasal interfaces in preterm infants extubated to NCPAP following IPPV for RDS or treated with NCPAP soon after birth.

Zbieranie i analiza danych

Data was extracted and analysed by the first three authors. Dichotomous results were analysed using the relative risk (RR), risk difference (RD) and number needed to treat (NNT).

Główne wyniki

1. Preterm infants being extubated to NCPAP following a period of IPPV for RDS:

Meta-analysis of the results from Davis 2001 and Roukema 1999a demonstrated that short binasal prongs are more effective at preventing re-intubation than single nasal or nasopharyngeal prongs [typical RR 0.59 (CI: 0.41, 0.85), typical RD -0.21 (CI: -0.35, -0.07), NNT 5 (CI: 3, 14)]. In one study comparing short binasal prong devices (Sun 1999), the re-intubation rate was significantly lower with the Infant Flow Driver than with the Medicorp prong [RR 0.33 (CI: 0.17, 0.67), RD -0.32 (CI: -0.49, -0.15), NNT 3 (CI: 2, 7)]. The other study comparing short binasal prong devices (Infant Flow Driver versus INCA prongs, Stefanescu 2003) demonstrated no significant difference in the re-intubation rate but did show a significant reduction in the total days in hospital in the Infant Flow Driver group [MD -12.60 (95% CI: -22.81, -2.39) days].

2. Preterm infants primarily treated with NCPAP soon after birth:

In the one trial identified, Mazzella 2001 found a significantly lower oxygen requirement and respiratory rate in those randomised to short binasal prongs when compared with CPAP delivered via nasopharyngeal prong. The requirement for intubation beyond 48 hours from randomisation was not assessed.

3. Studies randomising preterm infants to different NCPAP systems using broad inclusion criteria

The studies of Rego 2002 and Buettiker 2004 did not examine the primary outcomes of this review. Of the secondary outcomes, Rego 2002 demonstrated a significantly higher incidence of nasal hyperaemia with the use of the Argyle prong compared with Hudson prongs [RR 2.39 (95% CI: 1.27, 4.50), RD 0.28 (95% CI: 0.10, 0.46)].

One study comparing different techniques of pressure generation is awaiting further assessment as it is currently available in abstract form only.

Wnioski autorów

Short binasal prong devices are more effective than single prongs in reducing the rate of re-intubation. Although the Infant Flow Driver appears more effective than Medicorp prongs the most effective short binasal prong device remains to be determined. The improvement in respiratory parameters with short binasal prongs suggests they are more effective than nasopharyngeal CPAP in the treatment of early RDS. Further studies incorporating longer-term outcomes are required. Studies are also needed to determine the optimal pressure source for the delivery of NCPAP.

Cytowanie
De Paoli AG, Davis PG, Argus B, Jackson HD. Devices and pressure sources for administration of nasal continuous positive airway pressure (NCPAP) in preterm neonates. Cochrane Database of Systematic Reviews 2008, Issue 1. Art. No.: CD002977. DOI: 10.1002/14651858.CD002977.pub2.

Używamy plików cookie

Używamy niezbędnych plików cookie, aby nasza strona mogła działać. Chcielibyśmy również ustawić opcjonalne pliki cookie do analizy, aby pomóc nam udoskonalić tę stronę. Nie będziemy ustawiać opcjonalnych plików cookie, chyba że je włączysz. Użycie tego narzędzia spowoduje ustawienie pliku cookie na Twoim urządzeniu, aby zapamiętać Twoje preferencje. W każdej chwili możesz zmienić swoje preferencje dotyczące plików cookie, klikając na link "Ustawienia plików cookie" znajdujący się w stopce każdej strony.
Bardziej szczegółowe informacje na temat używanych przez nas plików cookie można znaleźć na naszej stronieStrona plików cookies

Zaakceptuj wszystkie
Skonfiguruj