Direkt zum Inhalt

Umbilical artery catheters in the newborn: effects of catheter design (end vs side hole)

Auch verfügbar in

Using umbilical artery catheters with the hole in the side instead of end leads to serious complications. The umbilical artery catheters (tubes) (UACs) commonly used in neonatal intensive care to monitor babies can sometimes cause them problems. They can be used with different accompanying doses of the drug heparin, and in different positions. The design of catheters can vary. The review found that catheters with a hole in the side instead of at the end resulted in one in three babies experiencing aortic thrombosis (a clot in the heart's major artery).

Hintergrund

Umbilical arterial catheters (UACs) are among the most commonly used monitoring methodologies in neonatal intensive care. Immediately after insertion of a UAC, local vascular compromise is frequently evident, usually in the form of blue or white toes, but occasionally with more severe and extensive ischemic manifestations. Aortic thrombi and renal ischemia have also been described. Some epidemiologic and case control studies have shown that the use of umbilical artery catheterisation is statistically associated with the later development of necrotizing enterocolitis.

Zielsetzungen

To determine whether the design of an umbilical arterial catheter influences the frequency of ischemic events, aortic thrombosis, intraventricular hemorrhage, mortality or necrotising enterocolitis in newborn infants.

Suchstrategie

Randomized and quasi-randomized controlled trials of umbilical catheterization use were obtained using the search methods of the Cochrane Neonatal Review Group. The Cochrane Library, MEDLINE (search via PubMed), CINAHL and EMBASE were searched from 1999 to 2009.

Auswahlkriterien

Randomized trials in newborn infants of any birthweight or gestation.
Comparison of end hole catheters with side hole catheters.
Clinically important end points such as ischemic events or aortic thrombosis.

Datensammlung und ‐analyse

There appears to be only a single trial which has addressed this issue (Wesstrom 1979).

Hauptergebnisse

End hole catheters are associated with a much decreased risk of aortic thrombosis compared to side hole catheters. RR = 0.27 (95% CI 0.11, 0.67)

Schlussfolgerungen der Autoren

Side hole catheters should be avoided for umbilical arterial catheterisation in the newborn.

Zitierung
Barrington KJ. Umbilical artery catheters in the newborn: effects of catheter design (end vs. side hole). Cochrane Database of Systematic Reviews 1999, Issue 1. Art. No.: CD000508. DOI: 10.1002/14651858.CD000508.

So verwenden wir Cookies

Wir verwenden notwendige Cookies, damit unsere Webseite funktioniert. Wir möchten auch optionale Cookies für Google Analytics setzen, um unsere Webseite zu verbessern. Solche optionalen Cookies setzen wir nur, wenn Sie dies zulassen. Wenn Sie dieses Programm aufrufen, wird ein Cookie auf Ihrem Gerät platziert, um Ihre Präferenzen zu speichern. Sie können Ihre Cookie-Einstellungen jederzeit ändern, indem Sie auf den Link "Cookie-Einstellungen" am Ende jeder Seite klicken.
Auf unserer Seite zu Cookies finden Sie weitere Informationen, wie diese Cookies funktionieren die Seite mit den Cookies.

Alle akzeptieren
Anpassen