The frameless IUD performs similarly to traditional IUDs but does not reduce bleeding and pain associated with standard IUDs.

Također dostupno na

Devices placed in the uterus are highly effective at preventing pregnancy. Traditional intrauterine devices (IUD) with plastic frames have side effects such as excessive bleeding and pain that were thought to be due to the frame. This review found that symptoms of bleeding and pain, and contraceptive efficacy were not improved with the frameless device. Trials are needed to see if the frameless IUDs could benefit women who have not had children.

Uvod

The frameless intrauterine device (IUD) dispenses with the frame in the classical IUD and holds the device in the uterus by anchoring one end of a nylon thread in the fundal myometrium, to which copper sleeves are attached.

Ciljevi

This review examines the hypothesis that the frameless IUD Gynefix reduces risk of expulsion and pregnancy, and the problems of bleeding and pain necessitating early removal.

Metode pretraživanja

We searched the Cochrane Controlled Trial Register (Cochrane Library Issue 2, 2004), MEDLINE, and Popline from 1980 to March 2004, and reference list of articles.

Kriteriji odabira

We selected for the review randomised trials that compared the frameless device to a classical framed device for contraception.

Prikupljanje podataka i obrada

Both authors extracted data independently. We contacted study author for additional data. We calculated rate ratios and rate differences for cumulative rates for each outcome at yearly intervals. We used the inverse variance-based method to combine trials, and tested the results for heterogeneity.

Glavni rezultati

Four trials were included in the review involving 5,939 women randomised to either a frameless device or TCu380, with data up to eight years for the largest, and with a total experience of 23,180 years. Apart from one small trial, nulliparous women were excluded from the trials. The two earlier trials used a prototype introducer and there was a higher expulsion rate at one year (relative risk 2.48, 95% confidence intervals 1.89 to 3.26). However, between two and six years in the large WHO trial the risk of pregnancy was lower with the frameless device (relative risk 0.53, 95% confidence intervals 0.32 to 0.91). In a recent trial using GyneFix with a new introducer early expulsions and pregnancies were not statistically different from the control device. Removals rates for excessive bleeding and/or pain were no different between the devices (relative risks 0.92, 95% confidence intervals 0.74 to 1.14, at one year and 1.13, 0.93 to 1.37, at six years). There was a tendency towards fewer removals for pain in early years but no difference at six years (relative risk 1.13, 95% confidence intervals 0.93 to 1.37).

Zaključak autora

There is insufficient data to show that problems of early expulsions have been overcome with the modified introducer used in GyneFix. Apart from that, the frameless device performs similarly to TCu380, and appears to have a lower pregnancy rate in later years, although the absolute difference is small.

Citat
O'Brien P, Marfleet CC. Frameless versus classical intrauterine device for contraception. Cochrane Database of Systematic Reviews 2005, Issue 1. Art. No.: CD003282. DOI: 10.1002/14651858.CD003282.pub2.

Naše korištenje kolačića

Koristimo nužne kolačiće kako bi naša web stranica radila. Željeli bismo postaviti i neobavezne analitičke kolačiće koji će nam pomoći da ju poboljšamo. Nećemo postaviti neobavezne kolačiće ako ih ne omogućite. Korištenjem ovog alata postavit će se kolačić na vaš uređaj, kako bi zapamtili vaše postavke. Svoje postavke kolačića možete promijeniti u bilo kojem trenutku klikom na vezu "Postavke kolačića" u podnožju svake stranice.