Closed tracheal suction systems versus open tracheal systems for mechanically ventilated adults

The comparison of open and closed suction systems shows them to have similar results in terms of safety and effectiveness.

Tracheal secretions in mechanically ventilated patients are removed using a catheter via the endotracheal tube. The suction catheter can be introduced by disconnecting the patient from the ventilator (open suction system) or by introducing the catheter into the ventilatory circuit (closed suction system). Although the literature reports several advantages for the closed suction system, the review did not show differences between the two systems in the main outcomes studied. These outcomes were ventilator-associated pneumonia and mortality. This review identified few trials of high methodological quality. Future research should be of higher quality, clarify issues related to the patient's condition and to technique, and provide nurse-related outcomes.

Wprowadzenie

Ventilator-associated pneumonia is a common complication in ventilated patients. Endotracheal suctioning is a procedure that may constitute a risk factor for ventilator-associated pneumonia. It can be performed with an open system or with a closed system. In view of suggested advantages being reported for the closed system, a systematic review comparing both techniques was warranted.

Cele

To compare the closed tracheal suction system and the open tracheal suction system in adults receiving mechanical ventilation for more than 24 hours.

Metody wyszukiwania badań

We searched CENTRAL (The Cochrane Library 2006, Issue 1) MEDLINE, CINAHL, EMBASE and LILACS from their inception to July 2006. We handsearched the bibliographies of relevant identified studies, and contacted authors and manufacturers.

Kryteria doboru badań

The review included randomized controlled trials comparing closed and open tracheal suction systems in adult patients who were ventilated for more than 24 hours.

Zbieranie i analiza danych

We included the relevant trials fitting the selection criteria. We assessed methodological quality using method of randomization, concealment of allocation, blinding of outcome assessment and completeness of follow up. Effect measures used for pooled analyses were relative risk (RR) for dichotomous data and weighted mean differences (WMD) for continuous data. We assessed heterogeneity prior to meta-analysis.

Główne wyniki

Of the 51 potentially eligible references, the review included 16 trials (1684 patients), many with methodological weaknesses. The two tracheal suction systems showed no differences in risk of ventilator-associated pneumonia (11 trials; RR 0.88; 95% CI 0.70 to 1.12), mortality (five trials; RR 1.02; 95% CI 0.84 to 1.23) or length of stay in intensive care units (two trials; WMD 0.44; 95% CI -0.92 to 1.80). The closed tracheal suction system produced higher bacterial colonization rates (five trials; RR 1.49; 95% CI 1.09 to 2.03).

Wnioski autorów

Results from 16 trials showed that suctioning with either closed or open tracheal suction systems did not have an effect on the risk of ventilator-associated pneumonia or mortality. More studies of high methodological quality are required, particularly to clarify the benefits and hazards of the closed tracheal suction system for different modes of ventilation and in different types of patients.

Cytowanie
Solà I, Benito S. Closed tracheal suction systems versus open tracheal suction systems for mechanically ventilated adult patients. Cochrane Database of Systematic Reviews 2007, Issue 4. Art. No.: CD004581. DOI: 10.1002/14651858.CD004581.pub2.

Używamy plików cookie

Używamy niezbędnych plików cookie, aby nasza strona mogła działać. Chcielibyśmy również ustawić opcjonalne pliki cookie do analizy, aby pomóc nam udoskonalić tę stronę. Nie będziemy ustawiać opcjonalnych plików cookie, chyba że je włączysz. Użycie tego narzędzia spowoduje ustawienie pliku cookie na Twoim urządzeniu, aby zapamiętać Twoje preferencje. W każdej chwili możesz zmienić swoje preferencje dotyczące plików cookie, klikając na link "Ustawienia plików cookie" znajdujący się w stopce każdej strony.